Terminal disassembly tool for Varian x-ray housing
We calculate the final price from your selected file. You review and accept it before payment.
Saved on this device.
Pinned tool for removing threaded plate in Varian x-ray tube housings
- Rating
- 0 / 5
- Likes
- 0
- Downloads
- 1
- Makes
- 0
- Views
- 10
- Reviews
- 0
Updated at source:
Model information
- Source
- Printables
- Category
- Medical Tools
- License
- Creative Commons — Public Domain
When disassembling a Varian x-ray housing, the terminal sleeve is held in place by a threaded plate with 3 pin holes. I made this tool to enable this plate to be easily spun out (and replaced, if you need). The housing I have disassembled is a "B-130H" but I imagine the terminals will be the same on many other units.
Use three M4x20 socket head screws and nuts to get the correct length protruding.
For general information: To disassemble these units after the outer end caps come off, one side has a bellows for oil expansion (leave this in place), the other side has electrical terminals. There's a very large circlip which holds the inner end caps on, but it's very easy to bend and remove. After that, the solid end with the electrical terminals can be removed just by pulling hard. It helps to thread the terminal block screws back in so you have something to grab.